HUMIDITY GENERATORS MODEL 2015 - TS
The Model 2015 – TS is the latest chamber from GEO in the Model 2000 series.
Features:
- Embedded Touch Screen
- Remote Login Capabilities
- Automatic Charting of Probe Data
- Custom Doors for Dataloggers and Chart Recorders
Notes
- Typical performance specification for the system’s internal chamber. Insert is required for higher accuracy.
- Specified low temp is 5°C, but typically temp is 15°C below ambient. For high accuracy, insulated door is recommended.
- Adjustment of system using standard calibrated by 2-pressure readings
- After chamber reaches and maintains temperature equilibrium.
- Laboratory certificates of Thunder Scientific control probe calibration (Accredited NVLAP 200582-0) provided with system.
- This accuracy stated represents a worst case scenario. Typically the accuracy is ±0.6% or better.
- To achieve low RH values, dessicant must be 50% unused or better. To achieve high humidity values, system must be warmed up.
- System software calculates dew point in real time to prevent condensation.